Description
Oxygen error: Oxygen output
concentration has been below
50% for 10 minutes.
Oxygen delivery error: A breath
has been recognized, but proper
oxygen delivery has not been
detected.
Battery empty, attach plug:
The concentrator has insufficient
battery power. The concentrator
will shut down and stop
producing oxygen.
Battery HOT : The battery has
exceeded temperature limit
while concentrator is running on
battery power. The concentrator
will shut down and stop
producing oxygen.
System HOT: Concentrator
temperature is too high. The
concentrator will shut down and
stop producing oxygen.
Sensor fail: The
concentrator's oxygen
sensor has malfunctioned.
System COLD : The system is
cold (<2˚C). The concentrator will
shut down and stop producing
oxygen.
Page 24 of 272
What To Do
If condition persists, switch to
your backup oxygen source
and contact your equipment
provider to arrange for
service.
If condition persists, switch to
backup oxygen source and
contact your equipment
provider to arrange for
service.
Attach an external power
supply or replace with a fully
charged battery. If the device
has turned off, press and hold
the power button to turn back
on.
If possible, move concentra-
tor to a cooler location, then
turn power off and back on.
Ensure air intake and outlet
vents have clear access and
particle filters are clean. If
condition persists, switch to
external power or a backup
source of oxygen and contact
your equipment provider.
Ensure air intake and outlet
vents have clear access and
particle filters are clean. If
condition persists, switch to
a backup source of oxygen
and contact your equipment
provider.
You may continue to use the
concentrator. If the condition
persists, contact your
equipment provider.
Move to a warmer
environment to allow the unit
to warm up before starting it.
If condition persists, switch to
a backup source of
oxygen and contact your
equipment provider.
96-13613-00-01 A