12. TECHNICAL AND PRODUCT SPECIFICATIONS
12.1 SPECIFICATIONS
Inogen Rove 6 Portable Oxygen Concentrator (Model # IO-501)
Mains Isolation
Dimensions with standard battery
Dimensions with extended battery 18.24 x 8.31 x 22.91 cm (7.18 x 3.27 x 9.02 in)
Weight with standard battery
Weight with extended battery
Nominal sound level
Warm up time
Oxygen concentration*
Inspiratory trigger pressure
sensitivity
Flow control settings
Maximum outlet pressure
AC Power
DC Power
Power Consumption
Battery type
Rechargeable battery
Minimum rated capacity
Battery charge current
Battery re-charge time
Operating temperature**
Operating humidity
Operating atmospheric pressure
Operating altitude**
Shipping and storage temperature -25 to 70˚C (-13 to 158˚F)
©2024 Inogen, Inc. All rights reserved.
Remove both the DC input cable from device as well as the
battery pack.
18.24 x 8.31 x 20.68 cm (7.18 x 3.27 x 8.14 in)
2.2kg (4.8 pounds)
2.6kg (5.8 pounds)
39 dBA typical at setting 2 (MDS-Hi)
Maximum system sound power of 62 dBA
Maximum system sound pressure of 54 dBA
Typical lowest alarm sound pressure of 62.3 dBA (Measured in the
carry bag)
Typical highest alarm sound pressure of 67.5 dBA (Measured in the
carry bag)
(Sound pressures measured at 1 meter per ISO 3744
2 minutes
90% + 6% and - 3% at all settings
<0.12 cm H20
Pulse dose setting 1,2,3,4,5,6
< 28.9 PSI (199 kPa)
100 to 240 VAC, 50 to 60 Hz
Autosensing 2.0 – 1.0A
13.5-15.0VDC,100W
Max voltage: 12.0 to 16.8 VDC (+ 0.5)
120 W max
Lithium ion
12.0 to 16.8 VDC (± 0.5V)
8 cells battery: 6400 mAh
16 cells battery: 2x6400 mAh
3A per 8 cells 1,5A/ Side (BA-500)
3A per 8 cells (BA-508)
4A per 16 cells 2,0A /Side
Standard (BA-500 & BA-508): up to 3 hours
Extended (BA-516): up to 4 hours
5 to 40˚C ( 41 to 104˚F)
15% to 90%, non-condensing
70 kPA to 106 kPA
0 to 3048 meters (0 to 10,000 ft)
Page 31 of 272
96-13613-00-01 A