• The settings of this device may not
correspond with the settings for devices
that provide a continuous flow oxygen.
• Use of this device at an altitude above 3,048
m (10,000 ft) or outside the temperature range
of 5 – 40°C (41 – 104° F) or a relative humidity
above 95% is expected to adversely affect
the flowrate and the percentage of oxygen
and consequently the quality of the oxygen
therapy. Use of this device immediately after
storage in temperatures beyond the
allowable operating range may adversely
affect operation of the device until the tem-
perature returns to the allowable operating
range. Wind or strong drafts can adversely
affect the accurate delivery of oxygen
therapy.
• If the device fails, it will cause a return to your
previous condition prior to starting oxygen
therapy. This state will be different for each
individual patient.
• If you are unable to communicate discomfort,
you may require additional monitoring and or
a distributed alarm system to convey the
information about the discomfort and or the
medical urgency to your responsible
caregiver to avoid harm.
4.2 CAUTION
Risk of minor injury or discomfort
• Use of this device has not been studied
in pediatric populations. Consult your
physician before using the product for
pediatric patients.
• Incompatible parts and accessories can
result in degraded performance or damage
and may void your warranty.
• The device is designed to provide a flow of
high purity oxygen. An advisory alert, "Oxygen
Low", will inform you if oxygen concentration
drops. If alarm persists, contact your equip-
ment provider.
• The oxygen flow setting must be determined
and recorded for each patient individually by
the prescriber, including the configuration of
the device, its parts and the accessories. It is
the patient responsibility to check with
provider to reassess settings of the therapy
for effectiveness.
• It is the responsibility of the patient to plan for
a back-up oxygen supply when traveling;
Inogen assumes no liability for any
disruptions in oxygen supply if a backup
source is not secured.
©2024 Inogen, Inc. All rights reserved.
• It is the responsibility of the patient to use
only parts and accessories mentioned in
these instructions for use. Parts and
accessories used by the patient not
recommended in these instructions for use
are at the sole responsibility for the patient.
Inogen assumes no liability for use of parts
and accessories not mentioned in these
instructions for use.
• It is the responsibility of the patient to
periodically check the battery and replace
as necessary per these instructions for
use. Inogen assumes no liability for persons
choosing not to adhere to manufacturer
recommendations.
• Do not modify the device. Incompatible parts
and accessories as a result of modifications
can degrade performance or cause damage
and may void your warranty unless indicated
or instructed to do so.
• Do not use this product in any way other than
described in the specifications and intended
use sections of this manual as it may lead to
product damage, loss of product function, or
personal injury.
• Do not obstruct air intake or exhaust when
operating the device. Blocking air circulation
or placing close to a heat source may lead
to internal heat buildup and shutdown or
damage to the concentrator. In the event of
changes to the performance of the device,
please refer to the troubleshooting section of
this document.
• Do not operate the device without the particle
filter in place. Particles drawn into the system
may damage the equipment.
• Do not wrap cords around power supply for
storage. Do not drive, drag or place objects
over cord. Doing so may lead to damaged
cords and a failure to provide power to the
concentrator.
• Do not use the DC power cord with a plug
splitter. This may cause overheating of the
DC power cord.
• Do not disassemble the power supply. This
may lead to component failure and/or safety
risk.
• Do not place anything in the device's power
port other than the supplied power supply.
If an extension cord is used, use an extension
cord that has an Underwriters Laboratory
(UL) Mark and a minimum wire thickness of 18
gauge. Do not connect any other devices to
the same extension cord.