• Battery Storage: Remove your battery from the device when it is not in use to avoid inadvertent
discharge. Store battery in a cool, dry place. Store with a charge of at least 40-50%. Batteries
should be charged up to full charge and discharged down to 0% at least once every 90 days to
maintain maximum lifetime. Avoid storing your device Battery in extreme temperatures, below
-20˚C (-4˚F ) or above 60˚C (140˚F ), for any amount of time.
• Battery Disposal: Batteries must only be placed in the collection containers for waste portable
batteries according to local governing ordinances in sorted municipal waste or by waste recycling
organizations. Batteries must be discharged, or precautions against short circuits must be taken in
the case of batteries that are not completely discharged (e.g. by isolating the poles with adhesive
tape). Lithium-ion batteries, like all rechargeable batteries, are recyclable and should never be
incinerated.Old batteries may have possible negative effects on the environment and human health,
and it is the obligation for every user to follow the local rules on the separate collection and recycling
of waste batteries.
10.7 DC POWER CABLE FUSE REPLACEMENT (RP-125)
The DC power cable contains a fuse. If the DC power cable is being used with a known good power
source and the device is not receiving power, the fuse may need to be replaced.
To replace the fuse:
1. Remove the tip by unscrewing the retainer. Use a tool if
necessary.
2. Remove the retainer, tip, and fuse.
3. The spring should remain inside the adapter housing.
4. If the spring is removed, replace the spring first before
inserting the replacement fuse.
5. Install a replacement fuse
6. Reassemble the tip.
7. Ensure the retainer ring is properly seated and tightened.
WARNING
• CHOKING HAZARD: small parts exposed when changing the fuse, keep away from small
children and pets.
• CRITICAL FUSE SIZING: incorrect fuse replacement size may result in fire or inadequate
equipment protection. Replace only with same type and rating of fuse.
• ELECTRICAL SHOCK: completely disconnect the cable before attempting to change the fuse.
• Do not hang any type of accessory or accessory bracket from plug or cable.
11. DEVICE REPAIR & DISPOSAL
11.1 REPAIR
Do not attempt to repair the device unless otherwise specified in these instructions for use.
Contact your equipment provider or Inogen for assistance.
11.2 DISPOSAL
Follow your local governing ordinances for disposal and recycling of the device, accessories and
packaging. All electronic devices are subject to the WEEE regulation and must be disposed according
to local governing ordinances in sorted municipal waste or by waste recycling organizations. The
battery contains lithium-ion cells and should be recycled. The battery must not be incinerated, see
section 10.6.
©2024 Inogen, Inc. All rights reserved.