high viscosity or in suspension). For further information, please refer to the package leaflet provided
by the drug manufacturer. - Use the nebuliser in the correct position, as upright as possible; do not tilt
the nebuliser beyond an angle of 30 degrees, in any direction, to prevent the drug from spilling into
the mouth or being overdispersed, reducing the effectiveness of the treatment. - Pay attention to the
indications provided with the drug and avoid using the devices with substances and dilutions other
than those recommended. - Only use the device in a dust-free environment, otherwise therapy may be
impaired. - Do not obstruct or insert objects into the filter and its housing in the unit.
- In the case of substances that are too dense, dilution with suitable saline solution may be necessary,
as prescribed by a doctor. - Only use original Flaem accessories or spare parts, no liability is accepted if
non-original parts or accessories are used.
Risk of infection : - We recommend personal use of the accessories to avoid any risk of infection.
- Follow hygienic preparation before each use. Ensure that tubing and accessories are not stored in the
vicinity of other accessories or devices for different therapies (e.g. infusions). - At the end of therapy,
do not leave the medicine inside the nebuliser and proceed with the hygienic preparation operations.
- If the nebuliser is used for several types of medicine, the residues must be completely removed. There-
fore, perform hygienic preparation after each inhalation, also to achieve the highest degree of hygiene
and to optimise the service life and operation of the device.
Risk of injury: - Do not place the device on a soft supporting surface such as a sofa, a bed or a tablecloth.
- Always operate it on a hard surface clear of any objects.
WARNINGS ON INTERFERENCE RISKS DURING USE IN DIAGNOSTIC INVESTIGATIONS.
This device is designed to meet the current requirements for electromagnetic compatibility. As far as EMC
requirements are concerned, electromedical devices require special care when being installed and used,
and are therefore required to be installed and/or used in accordance with the manufacturer's specifications.
Risk of potential electromagnetic interference with other devices. Mobile or portable RF radio and
telecommunication devices (mobile phones or wireless connections) may interfere with the operation of
electromedical devices. The device may be susceptible to electromagnetic interference in the presence of
other devices used for specific diagnosis or treatment. For more information visit www.flaemnuova.it.
CASE HISTORY OF FAULTS AND THEIR RESOLUTION. Before performing any operation, switch off
the appliance and disconnect it from the mains.
Problem
Cause
Power cable not correctly
Thedevice is not
plugged into the appliance
working
socket or mains power socket
The medicine was not inserted
into the nebuliser
The nebuliser has not been
assembled correctly.
The device does
The nebuliser is obstructed
not nebulise
or nebulises
poorly
The accessories are not properly
connected to the appliance
The pipe is bent or damaged
or twisted
The air filter is dirty
The device is
louder than
The filter is inserted
usual
If, after checking the conditions described above, the device still is not working properly, we recom-
mend that you contact your trusted dealer or an authorised FLAEM service centre nearest to you. You
can find a list of all Service Centres at http://www.flaemnuova.it/it/info/assistenza
Remedy
Plug the power cable correctly into the sockets
Pour the right amount of medicine into the
nebuliser
Disassemble and reassemble the nebuliser
correctly according to the connection diagram on
the cover.
Hygienic preparation of the nebuliser. Deposits
of medicine due to lack of hygienic preparation
of the nebuliser impair its efficiency and function.
Strictly follow the instructions in the HYGIENIC
PREPARATION chapter.
Check the correct connection between the air
inlet of the appliance and the accessories (see
connection diagram on cover).
Unwind the pipe and check it for crushing or
punctures. Replace it if necessary.
Replace the filter
Insert the filter properly into the housing