unit and let the unit run for 10 minutes to clean the nebulizer sheet.
7. After cleaning, absorb the water with an unused, clean cloth.
8. Wipe the surface of the main unit with 75% medical alcohol and
allow it to air dry, or use a clean soft cloth.
9. After completing all the above steps, store the main unit,
medication container and accessories in a dry, clean place.
Attention:
• Do not put the medication cup and accessories in boiling water for
disinfection, as it could become deformed. Do not put it in a
microwave to dry.
• The parts disinfected with disinfectant must be completely cleaned.
Otherwise, the residual disinfectant may cause worsening of
symptoms.
5.2 Replacing the medication container
The nebulizer sheet is a type of consumable. Generally, the nebulizer
life is about six months (at 20 minutes per use, three times a day). Life
depends on the application, medication and cleaning. If there is no or
little nebulization during use, replace the medication container in a
timely manner (please contact your local distributor for assistance).
5.3 Transportation and storage
Transportation and storage environment:
Temperature: -40°C ~ +55°C
Relative humidity: 5 % ~ 96 %
Air pressure: 500 hPa ~ 1060 hPa
Requirements for transportation and storage
- No corrosive gases and a well-ventilated room
- Keep the device out of the reach of children
- Do not store the device where it will be exposed to direct sunlight,
high temperatures, humidity, dust or water
- Avoid subjecting the device to tilt, vibration or shock.
- Transportation should be carried out in general means of
transportation or in accordance with contractual requirements.
EN