• Due to the different properties of the medical solutions, the device
may not switch off automatically when the medication has been
used up, so the user must switch the device off manually to protect
the nebulizer sheet from damage.
• Please hold the device steady during use and do not shake it
vigorously.
• There is an air hole on the lid of the medicine cup, do not cover it to
ensure normal atomization.
• The duration of each inhalation should not exceed 20 minutes. If you
experience any discomfort during use, please discontinue treatment
and consult a doctor.
4. Switch off the device: If you need to switch off the appliance during
use, please press the On/Off button. After 1 second, the indicator light
goes out and the appliance switches off.
Attention:
• To prevent the device from being operated accidentally (e.g. when
carried in a pocket), the product is designed to be switched on and
off after a long press of 1 second.
• After the end of use, it is normal for some medication to remain in
the medication container after the automatic switch-off.

4.3 Charging the lithium battery

Operating time of the battery:
(1) The indicator lights up orange when the battery is low, prompting
the user to charge the device.
(2) In normal use, the battery can work continuously for 1 hour after a
full charge.
Charging:
(1) Plug the charging cable into the port on the device.
(2) Plug the USB connection of the charging cable into an adapter and
plug it into a socket.
During the charging process, the indicator lights up blue and goes out
automatically when charging is complete.