Appendix I
Curve chart of equivalent volume particle diameter distribution:
The median particle diameter (D 0.50) is 1~4 µm. Error shall be within ±25 %
Appendix II Electromagnetic Compatibility (EMC)
Warning
◼
Don't near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for magnetic
resonance imaging, where the intensity of EM DISTURBANCES is high.
◼
Use of the NE-M01L adjacent to or stacked with other equipment should be avoided because it could
result in improper operation. If such use is necessary, NE-M01L and the other equipment should be
observed to verify that they are operating normall.
◼
Use of accessories, transducers and cables other than those specified or provided by the manufacturer
of NE-M01L could result in increased electromagnetic emissions or decreased electromagnetic
immunity of NE-M01L and result in improper operation.
◼
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the NE-M01L, including cables
specified by the manufacturer. Otherwise, degradation of the performance of NE-M01L could result.
Note:
IN-100 (NE-M01L) needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided below.
The basic performance: Nebulization rate: ≥0.25 mL/min.
When the device is disturbed, the data measured may fluctuate, please measure repeatedly or in another
environment to ensure its accuracy.
Other devices may affect this device even though they meet the requirements of CISPR.
The following cable types must be used to ensure that they comply with interference radiation and
immunity standards:
Name
Power cord
Table 1
Radiated RF EMISSIONS CISPR 11
Radiated RF EMISSIONS CISPR 11
Guidance and Declaration - Electromagnetic Emissions
Emissions test
Cable length(m)
1.0
Compliance
Group 1
Class A