EN
2. Overview
2.1 Function and use
Intended use:
The product is intended for the nebulization of liquid medications for
inhalation by the patient. The device can be used on children and
adults at home, in the hospital and in outpatient aftercare.
Indication: Nebulizer therapy
Patient group: Adults and children
Intended users: Healthcare professionals and patients under their
direction
Contraindications: The product must not be used together with
pentamidine.
2.2 Features
Power supply: DC 5 V or 3.7 V rechargeable lithium battery
Input power: <5 VA
Atomization rate in level I: ≥ 0.25 ml/min
Atomization rate in level II: ≥ 0.15 ml/min
Noise level: ≤50 dB
Equivalent volume-particle size distribution: The volume distribution
ratio of small atomized particles (diameter <5 µm) is at least 60%
Type of protection against electric shock: Class Ⅱ device, internally
powered device
Degree of protection against electric shock: Type BF applied part
Degree of protection against ingress of liquid: IP22
Note: Please purchase a medical-grade power adapter from a qualified
manufacturer (input: AC 100-240V, 50-60Hz; output: DC 5V, 1.0A).
2.3 Operating environment
Temperature: 5°C ~40°C
Air humidity: 15 %~90 %.
Air pressure: 700 hPa~l060 hPa
Caution: This product is not suitable for use in environments with
strong electromagnetic interference (e.g. various medium/high