Removing the Protection Guard
screws
Replacing the Protection Guard
Removing the Guide Rail
screws
Replacing the Guide Rail
Secotom-10
Instruction Manual
When cutting very large PCB's or other samples it might be
necessary to remove the protection guard to be able to cut through
the entire length of a sample in one operation.
Move the manual cutting table into the top position and engage
the securing rod into the locking position.
Remove the two screws holding the protection guard using a 4
mm Allen key. Keep the protection guard together with the two
screws in a safe place.
Release the securing rod and move the manual cutting table all
the way down.
Move the manual cutting table into the top position and engage
the securing rod into the locking position.
Place the protection guard in position and secure it with the 2
screws using a 4 mm Allen key.
Release the securing rod and move the manual cutting table all
the way down.
Always put the protection guard back on again
Always use safety goggles when cutting.
When cutting very large PCB's or other samples it might be
necessary to remove the guide rail to be able to cut through the
sample in the correct position.
Remove the two screws holding the guide rail using a 3 mm
Allen key. Keep the guide rail together with the two screws in a
safe place.
Place the guide rail in position and secure it with the 2 screws
using a 3 mm Allen key.
IMPORTANT
after cutting large samples.
33