CLEANING
1. Remove the sensor from the patient and disconnect it from the patient cable.
2. Clean the surface of the cable by wiping it with a 70% isopropyl alcohol pad.
3. Wipe all surfaces of the cable.
4. Saturate another cloth or gauze pad with sterile or distilled water and wipe all surfaces of the cable.
5. Dry the cable by wiping all surfaces with a clean cloth or dry gauze pad.
CAUTION
• Do not immerse the connector on the cable in any liquid solution.
• Do not sterilize by irradiation, steam, autoclave, or ethylene oxide.
• Do not clean with chemicals not approved above.
SPECIFICATIONS
The RD SET MP Series Patient Cable is indicated for use with the following sensors:
RD SET MP Patient Cable Compatibility:
When used with Masimo SET Technology:
Sensors
RD SET Adt
RD SET Pdt
RD SET Inf
RD SET Neo
RD SET NeoPt/NeoPt-500
RD SET Trauma (Adult)
RD SET Newborn Neonatal
RD SET Newborn Infant/
Pediatric
RD SET DCI®
RD SET DCI-P
RD SET TC-I
When used with Philips FAST-SpO
Sensors
RD SET Adt
RD SET Pdt
RD SET Inf
RD SET Neo
RD SET NeoPt
RD SET Trauma (Adult)
1
RD SET Newborn Neonatal
1
RD SET Newborn Infant/Pediatric
RD SET DCI
RD SET DCI-P
RD SET TC-I
Provides saturation and pulse rate values when used with Philips technology but does not operate in the specialty mode unless used with Masimo
1
technology.
A
accuracy is a statistical calculation of the difference between device measurements and reference measurements.
rms
Approximately two-thirds of the device measurements fell within ± A
RD SET MP Series Patient Cables and the above Masimo RD SET Sensors have been tested with Masimo SET technology and
Philips FAST-SpO
technology. The saturation accuracy of the neonatal sensors were validated on adult volunteers and 1%
2
was added to account for the properties of fetal hemoglobin.
Saturation Accuracy
Weight
Range
No Motion
> 30 kg
10–50 kg
3–20 kg
< 3 kg
> 40 kg
< 1 kg
> 30 kg
< 3 kg
3–10 kg
10–30 kg
> 30 kg
10–50 kg
> 30 kg
3.5%
Technology:
2
Weight Range
> 30 kg
10–50 kg
3–20 kg
< 3 kg
> 40 kg
< 1 kg
> 30 kg
< 3 kg
3–10 kg
1
10–30 kg
> 30 kg
10–50 kg
> 30 kg
(70–100% SpO
)
2
Motion
No Motion
2%
3%
2%
3%
2%
3%
3%
3%
2%
3%
3%
3%
2%
3%
3%
3%
2%
3%
2%
3%
2%
3%
2%
3%
N/A
Saturation Accuracy
(70–100% SpO
No Motion
of the reference measurements in a controlled study.
rms
4
Pulse Rate Accuracy
(25–240 bpm)
Motion
3 bpm
5 bpm
3 bpm
5 bpm
3 bpm
5 bpm
3 bpm
5 bpm
3 bpm
5 bpm
3 bpm
5 bpm
3 bpm
5 bpm
3 bpm
5 bpm
3 bpm
5 bpm
3 bpm
5 bpm
3 bpm
5 bpm
3 bpm
5 bpm
3 bpm
N/A
)
2
2%
2%
2%
3%
2%
3%
2%
3%
2%
2%
2%
2%
3.5%
Low Perfusion Accuracy
(70–100% SpO
)
2
Saturation
Pulse Rate
2%
3 bpm
2%
3 bpm
2%
3 bpm
3%
3 bpm
2%
3 bpm
3%
3 bpm
2%
3 bpm
3%
3 bpm
2%
3 bpm
2%
3 bpm
3%
3 bpm
3%
3 bpm
3.5%
3 bpm
Pulse Rate Accuracy
(25–240 bpm)
No Motion
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
3 bpm
9005E-eIFU-0618