RD SET™ MP Series
Patient Cables
Reusable
Prior to using this Cable, the user should read and understand the Operator's Manual for the Device, this Directions
for Use and the Directions for Use for the Sensor.
INDICATIONS
The RD SET™ MP Series patient cables and Masimo RD SET sensors are indicated for the continuous, noninvasive monitoring
of arterial saturation (SpO
DESCRIPTION
The RD SET MP Series patient cables are for use on Philips modules with Masimo SET® Technology and devices with Philips
FAST-SpO
Technology.
2
On Philips modules with Masimo SET Technology, RD SET MP Series patient cables are for use with RD SET sensors.
On devices with Philips FAST-SpO
Technology is not available on devices with Philips FAST-SpO
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET® oximetry or licensed to
use Masimo sensors.
WARNINGS, CAUTIONS AND NOTES
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and
discard if cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and
sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable may result in intermittent readings,
inaccurate results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting
either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize the
cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage
the electrical components, potentially leading to patient harm.
• Caution: Replace the cable when a replace cable or equivalent message is displayed consistently. Consult monitoring
device operator's manual for more information.
• Note: The cable is provided with X-Cal® technology to minimize the risk of inaccurate readings and unanticipated
loss of patient monitoring. The cable will provide up to 17,520 hours of patient monitoring time when connected to a
monitoring device that supports X-Cal technology. Replace the cable when the patient monitoring time is exhausted.
Some legacy monitoring devices do not distinguish between the cable and the sensor. In these devices, a replace sensor
or similar message could pertain either to the sensor or the cable.
INSTRUCTIONS
A) Attach the RD SET MP Series patient cable to the Device
1. Refer to Fig. 1. Orient the connector of the cable to mate with the patient cable connector on the Module.
Caution: Ensure you are connecting a RD SET MP patient cable to either a Masimo SET or Philips FAST-SpO
connector on the Module. There are a number of similar connectors with different colors and different mechanical
keying. Never force a patient cable connector into the module. Failure to use a RD SET MP patient cable may result in
damage to the module, inaccurate readings, or no readings.
Note: There is a difference in color shade between the RD SET MP Patient Cable and the Philips Module patient cable
connector. However, this is an acceptable configuration. Masimo SET performance is not available when connected to a
device with Philips FAST-SpO
B) Attaching the RD SET MP Series patient cable connector to an RD SET disposable sensor connector
1. Refer to Fig. 2. Orient the sensor connector to the patient cable connector as shown.
2. Refer to Fig. 3. Insert the sensor connector completely into the patient cable connector until it locks in place as shown.
C) Attaching the RD SET MP Series patient cable connector to the RD SET reusable sensor connector
1. Refer to Fig. 4. Orient the sensor connector to the patient cable connector, as shown.
2. Refer to Fig. 5. Insert the sensor connector completely into the patient cable connector. Close the sensor connector
cover over the patient cable connector until it locks in place.
D) Disconnecting the RD SET MP Series patient cable connector from the RD SET disposable sensor connector
1. Refer to Fig. 6. Pull firmly on the sensor connector to remove it from the patient cable.
E) Disconnecting the RD SET MP Series patient cable connector from the RD SET reusable sensor connector
1. Refer to Fig. 7. While holding the sides of patient cable connector, lift the sensor connector cover from the patient
cable connector.
2. Refer to Fig. 8. Pull firmly on the sensor connector to remove it from the patient cable.

DIRECTIONS FOR USE

Not made with natural rubber latex
LATEX
PCX-2108A
02/13
) and pulse rate for adult, pediatric, infant and neonatal patients.
2
Technology, RD SET MP Series patient cables are for use with RD SET sensors. Masimo SET
2
Technology.
2
Technology.
2
3
en
Non-Sterile
cable
2
9005E-eIFU-0618