Operation
8.11.2
Selecting Convertible Units and their
Decimal Places
Procedure
Press and hold down the [Toggle between weight
t
units] button.
Select one of the parameters "Unit 1" – "Unit 4" in the
t
submenu. To do so, press the [Confirm] button.
Select the desired unit. To do so, press the [Confirm]
t
button.
Specify the decimal places for the selected unit. To
t
do so, press the [Confirm] button once again.
Select the desired number of display digits. To do so,
t
press the [Confirm] button.
8.11.3

Running the "Statistics" Application

The "Statistics" application saves up to 99 weight values
and evaluates these statistically.
The following values are saved and exported:
— Number of components
— Mean value
— Standard deviation
— Variation coefficient
Sum of all values
— Lowest value (minimum)
— Highest value (maximum)
— Spread: Difference between maximum and minimum
Requirements
— A printer or a PC is connected and configured.
— The "STATIST." application is selected.
Procedure
Place a sample on the weighing pan.
t
To save the weight value: Press the [Confirm] button.
t
The position of the saved value is displayed, e.g. "NO
y
1".
The recording of the statistics starts.
y
Remove the sample on the weighing pan.
t
To save the next value:
t
Place a new sample on the weighing pan and
t
press the [Confirm] button.
To switch between the display of the current weight
t
value, the number of saved components, and the
calculated mean in the results display: Press the [Up]
or [Down] button.
To delete all saved values: Press the [Back] button.
t
The evaluation is printed and the active GLP print is
y
ended.
To print and exit the current statistics, and to delete
t
the saved values: Press the [GLP] button.
34
Entris
BCE Operating Instructions
®
8.12 Printing Weighing Result with
ID Marking
The device, the sample and a batch can be assigned an
ID number. The ID numbers are exported during GLP-
compliant printing.
Requirements
— The identity number is specified (see Kapitel „7.3.20
"INPUT" Menu", Seite 29).
— The printing of the lot ID line in the GLP printout is
activated in the menu (see Kapitel „7.3.20 "INPUT"
Menu", Seite 29).
— The "SPL. ID" output is activated in the menu.
— The GLP-compliant printout is activated (see
Kapitel „7.3.7 "DATA.OUT." / "PRNT.PAR." Menu",
Seite 24).
Procedure
Start the printout. To do so, press the [Print] button.
t
The GLP header is printed with the ID marking of the
y
LOT ID set in the menu and the current weight value.
The [GLP] button appears in the operating display.
y
Confirm the [PRINT] button.
t
The SPL. ID and the current weight value are
y
exported.
Confirm the [PRINT] button.
t
The SPL. ID and the current weight value are
y
exported.
To exit the GLP printout: Press the [GLP] button.
t
The GLP footer is printed.