8.8
Printing Results of the
Calibration and Adjustment
Process
The results of the calibration and adjustment process
can be printed in a GLP printout.
Procedure
Set the GLP printout in the menu (setting
t
parameters, see Kapitel 7.3.7, Seite 24).
Calibrate the device.
t
Once the calibration is complete: The printing
y
process starts.
8.9

Weighing

Requirements
— The device has been leveled.
— The device is calibrated and adjusted.
NOTICE
Chemicals may damage the device or accessories!
Chemicals can attack the device or the connected
accessories internally and externally. This may damage
the device and accessories.
Use appropriate containers when weighing
t
chemicals.
Procedure
Zero the device. In order to do so, press the [Zero]
t
button.
If a container is being used for the sample:
t
Tare the device. In order to do so, press the [Tare]
t
button.
Place the sample in the container.
t
If no container is used for the sample: Place the
t
sample on the weighing pan.
The weight value is displayed depending on the
y
selected application program.
8.10 Setting or Changing an
Application
Procedure
In the "APPLIC." menu, select an application, e.g.
t
"STATIST."
Press the [Confirm] button.
t
Exit the menu.
t
8.11
Running Applications
(Examples)
8.11.1
Executing the "Toggle Between Weight
Units" Function
The "Toggle between weight units" function enables a
switch between a maximum of four different units. The
selected basic unit is displayed every time the device is
started (see "UNIT", Kapitel „7.3.1 "SETUP" / "BALANCE"
Menu", Seite 19). The units can be set during the
weighing process and the decimal places can be
adjusted.
Requirements
The "Toggle between weight units" function is activated
(see Kapitel „7.3.9 "APPLIC." / "WEIGH" Menu",
Seite 25).
The weight value must be valid.
Procedure
To switch the displayed weight unit during weighing
t
or before an application: Press the [Toggle between
weight units] button until the desired unit is
displayed.
The current weight value is displayed in the selected
y
unit.
Entris
BCE Operating Instructions
®
Operation
33