Warning message
Fault
SYS.ERR.
The system data
is faulty.
ERR 02
The device cannot
be calibrated
when starting the
calibration
function due to a
zero point error.
ERR 10
Taring is not
possible.
ERR 11
The weight value
cannot be saved in
the tare memory.
Cause
A memory error exists
in the control module.
The device was not
zeroed before
calibration.
The device is loaded.
The device cannot be
manually tared
because an
application program
has the tare memory
reserved.
The weight value is
negative or "zero".
Remedy
Switch the device off and
on again.
If the problem occurs
again: Contact Sartorius
Service.
Zero the device.
Check the preload and set
if necessary.
Remove the sample from
the weighing pan.
To release the tare
memory: Exit the
application program
with the [Back] button.
Check the sample being
weighed.
Zero the device before
placing the sample on the
balance if necessary.
Entris
BCE Operating Instructions
®
Malfunctions
Chapter, page
37