Display/error
Error Message
Mounting unit not connected
PT100 amplifier uncalibrated
Pressure Error
Mounting pressure too high
Error, hydraulic-pump overloaded
Heating Error
Insufficient Cooling
Top closure not unscrewed
44
ProntoPress-10
Instruction Manual

6. Trouble-Shooting

Cause
The plug for the mounting unit is not
connected properly. (Appears when
START
is pressed).
s
The temperature measurement
mechanism is uncalibrated or there
is an error in calibration.
Appears during start-up mode
together with a long beep.
Defect in the pressure system.
Appears, when the pre-set pressure
is not reached within 25 seconds.
Formation of gas in the mounting
cylinder.
Appears, if the actual force should
exceed 65 kN.
The motor has been overloaded.
Appears, either when pressing: RAM
UP/DOWN
or during the process.
O N
The plug for the mounting unit is not
correctly connected.
Appears, when the pre-set
temperature is not reached within 6
minutes.
Defect in the heating system.
Mains tap supplying cooling water is
either closed or not opened
sufficiently.
Filter at the water inlet is blocked.
The water is too hot.
Appears, when the cooling is
inadequate.
Insufficient water within the
Recirculation Cooling Unit. The water
is too hot.
Appears, when the cooling is
inadequate.
Defect in the cooling system.
The top closure has not been
unscrewed completely after a
mounting.
Appears when RAM UP
after a mounting.
Action
Insert the plug firmly. If message still
appears, call a Struers service
technician.
Call a Struers service technician.
Call a Struers service technician.
Cool the mounting unit down
completely.
Move the lower ram down.
Open the top closure.
Let the motor cool down and start the
mounting press again.
If it does not help:
Call a Struers service technician.
Re-connect the plug.
Call a Struers service technician.
Open the tap.
Clean the filter.
Fill to the correct water level. See the
section: Maintenance.
Call a Struers service technician.
Turn the top closure until it is
completely free of the threads. If this
has already been done, the problem
is that the closure cannot spring free.
O is pressed
Hold the swivel arm down with one
finger and press RAM UP
again.
O