Cooling of Cylinder

ProntoPress-10
Instruction Manual
The mounting cylinder can be cooled without pressure on. This
can be performed with the top closure removed or in place. This
feature can be used for supplementary cooling, if, for example, an
ongoing mounting process should be stopped.
Go to the Select Method Group Menu:
SELECT METHOD GROUP
1. STRUERS
3. GROUP NUMBER 3
F1:CONFIG.
F3:MAN. FUNC.
If the heading in the display is different, press Esc until the
Select Method Group Menu appears.
Press F3 to activate the manual
F3
functions.
D
State: OFF
Pre-set Temperature:
Actual Temperature:
F1:START HEATING
D
mO
Press MENU UP/DOWN
mN
come to the display for cooling:
D
State: OFF
Cooling Time:
Cooling Rate:
F1:START COOLING
Adjust the cooling time and cooling rate to the desired values.
See Method Edit.
If the cooling time is set to 0:00, the cooling time will
automatically count from the maximum value: 15:00, when the
cooling is started.
2. GROUP NUMBER 2
4. GROUP NUMBER 4
F4:RENAME
HEATING OF CYLINDER
50 °C
25 °C
mO N
to
COOLING OF CYLINDER
4:30
HIGH
39