Verfügbare Sprachen

Display/error
Machine Problems
Insufficient compression.
Insufficient heating.
Insufficient cooling.
Cooling water drips underneath the
machine.
The top closure will not start
threading on the mounting cylinder.
46
ProntoPress-10
Instruction Manual
Cause
Incorrect setting of force/pressure.
Incorrect configuration of cylinder
diameter.
Incorrect unit for force or pressure.
Incorrect setting of time for
preheating or heating.
Incorrect temperature unit set.
Incorrect setting for cooling time.
Incorrect setting for cooling rate.
Incorrect Temperature unit set.
Mains tap supplying cooling water is
either closed or not opened
sufficiently.
Filter at the water inlet is blocked.
The water is too hot.
Insufficient water within the
Recirculation Cooling Unit. The water
is too hot.
Defect in the cooling system.
The quick coupling is not mounted
correctly.
The top closure is not mounted
correctly.
The upper ram is too hot.
Cured resin in the upper part of
mounting cylinder.
Cured resin on the cylindrical surface
of the upper ram.
The top closure has been dropped
on the floor/table, causing a bulge in
the edge of the ram.
The swivel arm has become askew.
Action
Set the correct parameter. The force
should be min. 5 kN/1000 Lbf/500kp.
Check the configuration.
Set the correct parameter.
Check the configuration.
Set the correct parameter.
Check the configuration.
Open the tap.
Clean the filter.
Fill to the correct water level. See the
section: Maintenance.
Call a Struers service technician.
Remove the cover from the mounting
unit and check the quick coupling
connections.
Press the top closure straight down,
turning it counter-clockwise until you
hear a click.
Turn the top closure clockwise.
Let the top closure and the upper
ram cool down.
Reduce the mounting temperature.
Clean the mounting cylinder with a
brass wire brush.
Clean the ram with the scraper
supplied.
Replace the upper ram.
Call a Struers service technician.
In the meantime the mounting press
can be used without the swivel arm.