7. Faults
Display
Cause
Measured value above allowable
Err.1
range
Sensor defect
Measured value below allowable
Err.2
range
Sensor defect
Scale range exceeded
Err.3
Under the scale range
Err.4
System error
Err.7
Measured value could not be
Er.11
calculated
Calculation range exceeded
WIKA operating instructions, model CPH6200
Measures
Check: Is the pressure over the
permissible measuring range of the
sensor?
Measured value too high!
Reduce pressure
Check the pressure measuring range
of the sensor and, if necessary,
replace with a suitable sensor with a
higher measuring range.
Send in for repair
Check: Is the pressure under the
permissible measuring range of the
sensor?
Measured value too low!
Reduce pressure
Check the pressure measuring range
of the sensor and, if necessary,
replace with a suitable sensor with a
lower measuring range.
Send in for repair
Check: Is the value over 9999?
Value is too high!
Reduce value
Check: Is the value below -2000
(tare?)
Value is too low!
Increase the value
Send in for repair
Choose different unit
Choose different unit