6. Commissioning, operation
Menu
Para-
Values
meter
[MENU]
Set Calibration: Adjustment of sensors
SEt
Sensor dep.,
CAL
OFS.1
e.g. -5.00 ...
+5.00 mbar
oFF
e.g. -5.00 ...
SCL.1
+5.00 mbar
oFF
Sensor dep.,
OFS.2
e.g. -5.00 ...
+5.00 mbar
oFF
e.g. -5.00 ...
SCL.2
+5.00 mbar
oFF
Set Alarm: Alarm function configuration
SEt
AL.
AL.
on
[1,2,DIF]
no.So
oFF
Min-Range ...
AL.Lo
... AL.Hi
[1,2,DIF]
AL.Lo ...
AL.Hi
... Max-Range
[1,2,DIF]
Set Logger: Logger function configuration
SEt
Lo66
Func
CYCL
Stor
oFF
1 ... 3600
CYCL
on / oFF
Lo.Po
Set clock: Setting of real-time clock
SEt
HH:MM
CLOC
CLOC
YYYY
YEAr
TT.MM
dAtE
1) If there is data in the logger memory, these menu items cannot be accessed.
If these are to be changed, the data must first be deleted, see chapter 6.5 "Data logger".
2) This menu can only be selected if an appropriate sensor is connected to connection 1. When using a second corresponding
sensor on connection 2 then the settings are adopted.
3) If the automatic recording is activated, these menu items cannot be accessed. If these should be changed, the cyclic data logger
must first be stopped, see chapter 6.5 "Data logger".
4) If an alarm function limit is crossed (over or under), this is signalled by a "hooting" and a beeping.
WIKA operating instructions, model CPH6200
Meaning
or
The zero point of sensor 1 will be displaced by this value to
compensate for deviations of the probe or the measuring
instrument.
Zero point offset is deactivated (= 0.000)
The measuring slope of sensor 1 will be changed by this
factor [%] to compensate for deviations of the probe or the
measuring instrument.
Slope correction factor deactivated (= 0.000)
The zero point of sensor 2 will be displaced by this value to
compensate for deviations of the probe or the measuring
instrument.
Zero point offset is deactivated (= 0.000)
The measuring slope of sensor 2 will be changed by this
factor [%] to compensate for deviations of the probe or the
measuring instrument.
Slope correction factor deactivated (= 0.000)
Alarm sensor is on; is indicated acoustically
Alarm sensor is on; is indicated acoustically
No alarm function
Min. alarm limit (not when AL.oFF, sensor min. is the lower
display range limit of the connected sensor)
Max. alarm limit (not when AL.oFF, sensor max. is the upper
display range limit of the connected sensor)
Cyclic: Logger function "cyclic logger"
Store: Logger function "individual value logger"
No logger function
Cycle time for cyclic logger [seconds]
Low-power logger with very low current supply (only with
cyclic logger and slow measurement)
Setting the time hours:minutes
Setting the year
Setting the date day.month
3)
4)
1)
EN
21