Manual Cleaning without Cleaning Tools
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
≤ 38°C, ≥ 3.5 L/min, ≥ 15 sec.
Water quality: Drinking water
_ Cleaning
1
Prepare cleaning solution in the cleaning tank under the
following conditions (prepare the cleaning solution according to
the instructions of the detergent manufacturer).
Temperature
Water quality
Distilled water
Deionized water
45°C
Demineralized
water
* Detergent used for validation:
neodisher MediZym (DR. WEIGERT)
2
Immerse the product in the cleaning tank for 20 - 30 min.
_ Drying
1
Remove moisture inside and on the surface of the product using
a dry cloth, air gun, etc.
Detergent
Neutral enzymatic
detergent for medical
purposes (pH 7.0 - 8.0)
25
Cleaning / Checking the Irrigation Nozzle
3
Thoroughly rinse the dirt released in immersion as well as the
detergent under running tap water.
≤ 38°C, ≥ 3.5 L/min, ≥ 10 sec.
Water quality: Drinking water
4
Rinse the product surface with distilled, deionized or
demineralized water.
2
Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Checking (see "9 – 2 Checking").

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis