Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
7 – 1 – 1 Manual Cleaning
Manual Cleaning using the Cleaning Tools
Use the Cleaning Tools according to the following system figure.
Attachment
Slim Attachment Hub
Slim Tube
Attachment
Attachment
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
For products with moving parts, perform cleaning while moving all parts (see "2 Moving Parts of Each Product").
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water
For Slim Attachment Hub, perform this step with the Slim Tube removed.
_ Cleaning
1
Turn the Twist Collet to the closed position (if applicable).
2
Mount the Cleaning Adaptor.
Pull the Cleaning Adaptor to make sure that it is securely
connected.
Cleaning Adaptor SM
Cleaning Adaptor 1T
Cleaning Adaptor HM
*1
Cleaning Adaptor WPD
*2
3
4
11
Cleaning
Cleaning Adaptor G
*1: With O-ring (Order Code: D0312074080)
*2: With O-ring (Order Code: D0312150150)
Screw the disposable luer-lock syringe (hereafter referred to as
syringe) into each Cleaning Adaptor of the Attachment.
Prepare cleaning solution in the cleaning tank under the
following conditions (prepare the cleaning solution according to
the instructions of the detergent manufacturer).
Temperature
Water quality
Distilled water
Deionized water
45°C
Demineralized
water
* Detergent used for validation:
neodisher MediZym (DR. WEIGERT)
Drying
Luer-lock syringe
Air Gun
Detergent
Neutral enzymatic
detergent for medical
purposes (pH 7.0 - 8.0)