Herunterladen Inhalt Inhalt Diese Seite drucken

NSK Primado2 Bedienungsanleitung Seite 25

Vorschau ausblenden Andere Handbücher für Primado2:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 32
Cleaning / Checking the Irrigation Nozzle
3
Prepare cleaning solution in the cleaning tank under the
following conditions (prepare the cleaning solution according to
the instructions of the detergent manufacturer).
Temperature
Water quality
Distilled water
Deionized water
45°C
Demineralized
water
* Detergent used for validation:
neodisher MediZym (DR. WEIGERT)
4
With the product immersed in the cleaning tank, pull the syringe
plunger to suck and discharge the cleaning solution. Perform
this step 5 times to retain the cleaning solution inside the
product.
Immerse the product in the cleaning tank for 20 - 30 min.
5
Push and pull the syringe plunger to discharge the cleaning
solution remaining in the product.
_ Drying
1
Remove moisture inside and on the surface of the product using
a dry cloth, air gun, etc.
Detergent
Neutral enzymatic
detergent for medical
purposes (pH 7.0 - 8.0)
x5
24
6
Thoroughly rinse the dirt released in immersion as well as the
detergent under running tap water.
≤ 38°C, ≥ 3.5 L/min, ≥ 10 sec.
Water quality: Drinking water
7
Put distilled, deionized or demineralized water into the syringe,
screw the syringe into the product, and discharge water from
the product tip.
Perform this step 5 times.
8
Remove the Cleaning Tube 1.5/L25 and the Cleaning Adaptor IN
from the Irrigation Nozzle.
9
Rinse the product surface with distilled, deionized or
demineralized water.
2
Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Checking (see "9 – 2 Checking").
x5

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis