Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
5
Place the Cleaning Holder in the cleaning basket and set the cleaning basket in the washer-disinfector.
6
Mount the Cleaning Tube 6.0/L650 to the washer-disinfector.
If necessary, insert Cleaning Adaptor Luer F, etc.
_ Cleaning and Disinfection
Perform cleaning and thermal disinfection using the washer-disinfector under the following conditions:
Temperature Time
Pre-cleaning
< 25°C
Cleaning
40 - 60°C
Rinsing
> 10°C
Thermal
90 - 93°C
disinfection
Drying
Set the washer-disinfector according to the instruction of the manufacturer.
* Detergent used for validation:
Neutral enzymatic detergent: neodisher MediZym (DR. WEIGERT)
Alkaline detergent: neodisher SeptoClean (DR. WEIGERT)
_ Drying
1
Screw the Cleaning Adaptor G into each Cleaning Adaptor.
2
Push the air gun into the Cleaning Adaptor G and remove
moisture inside the product.
Allowable air pressure
Purging Time
Water quality
3 min
Tap water
≥ 5 min
Distilled water
Deionized water
≥ 1 min
Demineralized water
5 min
< 500 kPa
≥ 5 sec.
Detergent
-
Neutral enzymatic detergent for medical purposes
(pH 7.0 - 8.0), or alkaline detergent for medical
purposes (pH ≤12)
-
-
3
Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
4
Remove the Cleaning Adaptor from the Attachment by
depressing the Release button.
5
Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Lubrication (see "7 – 2 Lubrication")