6. Saving, retrieving and deleting results

• The results of each successful measurement are stored
together with date and time. With more than 60 items of
measured data, the earliest items of data measured are lost.
• Select the desired user memory by pressing the MEM button
and then the + button. Press the MEM button again to dis-
play the average of all saved measurements in the user mem-
ory. Press the MEM button again to display the average day
measurement for the last 7 days. (day: 5.00 a.m. – 9.00 a.m.,
display ). Press the MEM button again to display the aver-
age night measurement for the last 7 days (night: 5.00 p.m. –
9.00 p.m., display ). If you continue pressing the MEM but-
ton, the latest individual results are displayed with date and
time.
• To clear the memory, press the MEM button, then the display
shows
Press the + button to select the user memory and
confirm by pressing MEM again. Press and hold the + and
keys simultaneously for 5 seconds (the display will show
).
• If you wish to change the user memory, see "Select mem-
ory".

7. Error messages / troubleshooting

In case of faults, the
message appears in the display.
Error messages can occur when
1. inflation pressure is higher than 300 mmHg,
2. blood pressure is unusually high or low,
3. you move or talk during the measurement (in addition to
the cardiac arrhythmia icon
4. the cuff tube is not inserted properly,
5. the cuff is inflated for longer than 25 seconds,
,
,
appears in the display),
,
.
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
not move or talk. Re-insert the batteries if necessary, or else
replace them.

8. Cleaning and Care

• Clean your blood pressure monitor carefully using a slightly
damp cloth only.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise liquid
can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.

9. Technical details

Model no.
Measurement
method
Measurement range
Display accuracy
Measurement
inaccuracy
,
Memory
Dimensions
17
BC 32
Oscillometric, non-invasive blood pres-
sure measurement on the wrist
Cuff pressure 0 – 300 mmHg,
systolic 40 – 280 mmHg,
diastolic 40 – 280 mmHg,
Pulse 40 –199 beats/minute
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 72 mm x W 69 mm x H 29.2 mm