BD Veritor™ InfoWiFi Instructions
Proximity Fields From RF Wireless Communications Equipment
Test
Band
a
frequency
(MHz)
(MHz)
385
380 – 390
450
430 – 470
710
704 – 787
745
780
810
800 – 960
870
930
1720
1700 – 1990
1845
1970
2450
2400 – 2570
5240
5100 – 5800
5500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is
permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it
does not represent actual modulation, it would be worst case.
28
Service
a
Modulation
TETRA 400
Pulse
modulation
18 Hz
GMRS 460,
FM
c
FRS 460
± 5 kHz
deviation
1 kHz sine
LTE Band 13, 17
Pulse
modulation
217 Hz
GSM 800/900,
Pulse
TETRA 800,
modulation
iDEN 820,
18 Hz
CDMA 850,
LTE Band 5
GSM 1800;
Pulse
CDMA 1900;
modulation
GSM 1900;
217 Hz
DECT;
LTE Band 1, 3,
4, 25; UMTS
Bluetooth,
Pulse
WLAN,
modulation
802.11 b/g/n,
217 Hz
RFID 2450,
LTE Band 7
WLAN 802.11 a/
Pulse
n
modulation
217 Hz
Maximum
Distance
b
power
(m)
(W)
1.8
0.3
b
2
0.3
0.2
0.3
b
2
0.3
b
2
0.3
b
2
0.3
b
0.2
0.3
b
TEST
LEVEL
IMMUNITY
(V/m)
27
28
9
28
28
28
9