Holder for the external canister "Bag"

Clean the surfaces of the holder for the external canister `Bag´ regularly and
disinfect them daily.
Cleaning
The holder for external canister "Bag" can be wiped daily in course of the cleaning of the device in
even if it is fitted, with a damp, lint-free cloth until no impurities are visible.
Manual Disinfection (optional)
A wipe disinfection can be done if internal hygiene requirements provide this.
Use disinfectants for wipe disinfection according to section 5.1 and observe the manufacturer´s
instructions.
Replace the holder for external canister "Bag" if the patient changes.
5.2.5

Suction tube

Cleaning
1. Rinse the suction tube under running water.
2. Wipe up the suction tube with a lint-free cloth.
3. Let the suction tube dry.
Manual Disinfection (optional)
1. Immerse suction tube in the disinfectant solution. Use disinfectants for immersion
disinfection according to section 5.1 and observe the manufacturer´s instructions.
2. Rinse the suction tube thoroughly and let it dry.
Replace the suction tube if the patient changes.
If the suction tube is reused on one patient, replace the suction tube weekly at the latest.
In the event of heavy contamination with sticky or encrusted secretion, the suction tube must be
replaced immediately.
5.2.6

Adapter for the disposable secretion canister system

Cleaning
1. Rinse the adapter for the disposable secretion system under running water.
2. Wipe up the adapter for the disposable secretion system with a lint-free cloth.
3. Let the adapter for the disposable secretion system dry.
Manual Disinfection (optional)
1. Immerse adapter for the disposable secretion system in the disinfectant solution. Use
disinfectants for immersion disinfection according to section 5.1 and observe the
manufacturer´s instructions.
2. Rinse the adapter for the disposable secretion system thoroughly and let it dry.
Replace the adapter for the disposable secretion system if the patient changes.