4.2.3
Connecting the disposable secretion canister system
Malfunction due to collapsing disposable liner "OneWay"
A leak in the external canister „Bag" or at the lid of the disposable liner
"OneWay" may cause air to flow into the external canister "Bag". This may
lead to the collapse of the disposable liner "OneWay".
• Ensure that the lid of the disposable liner „OneWay" is firmly connected to
the external canister „Bag".
• Ensure that all connections are firmly attached and properly connected.
• Ensure that the "Bag" external canister is undamaged and the T-piece is
firmly attached.
The original ASSKEA disposable secretion canister system consists of the external canister „Bag",
the holder for the external canister „Bag", the disposable liner „OneWay", the adapter for the
disposable secretion canister system and the suction tube.
10. Make sure that the adapter for the disposable secretion canister system is inserted
correctly, otherwise the holder for the external canister "Bag" cannot be inserted.
Page 64 of 84
Connection designation
A
Vacuum connection
B
Patient connection
Fig. 15