The reusable secretion canister system is for reuse on one patient.

3.6 Disposable secretion canister system for ASSKEA proVisio

secretion aspirator
The disposable secretion canister system consists of the external canister "Bag", the holder for
the external canister "Bag", the disposable liner "OneWay" and the suction tube. The disposable
liner "OneWay" has an internal bacterial filter, carbon filter, and solidifier. The self-sealing
bacterial filter integrated in the disposable liner "OneWay" is effective against microorganisms. It
prevents overflow in the event of an operational error. If the liquid reaches this filter, aspiration is
no longer possible and the vacuum setting buttons on the control panel illuminate continuously
red. The aspiration process is discontinued. The disposable liner "OneWay" must be replaced.
The activated carbon filter in the disposable liner "OneWay" reduces the spread of odor.
Solidifier:
Disposable liners "OneWay" that are filled with aspirate can be transported and disposed of in a
leak-proof manner with the aid of the solidifier. The aspirate solidifies after an average gelling
time of 2 to 5 minutes (depending on the consistency of the secretion), irrespective of the
aspiration intervals.
The disposable liner "OneWay" and the suction tube are for reuse on one patient.
Replace the disposable liner "OneWay" and suction tube according to the instruction in
section 4.3.3 and 5.2.5.

3.7 Information on the battery

The charge state of the battery is indicated by the corresponding LED status indicator (for
explanation, see section 4.3.1).
Prior to the initial startup of the ASSKEA proVisio
recommended to charge the battery completely and to repeat this after the initial uses.
Page 58 of 84
A
Lid reusable secretion canister
B
Overflow protection with ball
Fig. 4
®
M28 secretion aspirator, it is urgently
®