Malfunction
continuously
No suction
illuminate red
selected suction
setting
and
illuminate green
Device
switches off.
blinks red
Contact ASSKEA GmbH or your service partner if a malfunction cannot be
corrected by the described measures.
Page 78 of 84
Display
Probable cause
Ball of the overflow
protection blocks the
aspiration lid inlet
(reusable secretion
canister system)
disposable liner
„OneWay" is full.
Bacterial filter is blocked.
Suction tube is blocked
Adapter for disposable
secretion canister system
or bacterial filter is not
inserted
Internal error
Remedy
Check fluid level in the
reusable secretion canister,
empty the reusable
secretion canister.
Replace the disposable
liner „OneWay".
Replace the disposable
liner „OneWay" or the
bacterial filter.
Clean or change the
suction tube
Inserting adapter for
disposable secretion
canister system or bacterial
filter
Please contact service!