6 Troubleshooting

Malfunction
Display
do not illuminate
Device does
not start.
blinks red
blinks red
blinks red
and
Device does
not charge.
illuminates green
blinks red
illuminates green /
orange
and / or
illuminates green
Insufficient
suction
selected suction
setting
and
illuminate green
Probable cause
Battery is discharged
Battery is discharged and
impermissible input
voltage
Internal error
Battery temperature too
high for charging
Battery is defective if
charging is not possible
when the device has
cooled down.
Impermissible input
voltage
Leak in the suction line or
on the canister
Service life of vacuum
generator expired
Bacterial filter is blocked
Remedy
Connect the power supply
unit for charging or
operating.
Connect the correct power
supply unit, check the
power supply.
Please contact service!
No action required,
automatic resumption of
charging when battery has
cooled down
Please contact service!
Connect the correct power
supply unit, check the
power supply.
Check the lid of the
disposable liner „OneWay"
or the reusable secretion
canister lid and tube
connections for tight fit and
no leaks.
Please contact service!
Replace the disposable
liner "OneWay" or the
bacterial filter
Page 77 of 84