Operation of Conform Cannula Handpiece

By pressing the foot switch "Motor", the lifting movement of the Conform Cannula
Handpiece is actuated. The motor is running with a constant speed of 10,000 rpm.
The liposuction can be intermitted by pressing the vent button on the Conform Can-
nula Handpiece. By doing so the suction tube is ventilated and blockages of the tube
can be dissolved.
7.6

Emptying secretion jar

The jar lid of the secretion jar is equipped with an overflow protection system to prevent the vacuum system from
being flooded by secretion fluids at high filling levels of the secretion jar. Therefor a float gauge is responsible.
At high filling levels of the secretion jar and the resulting locking of the overflow protection system the secretion jar
has to be emptied or replaced by another secretion jar.
1. Swich off suction pump.
2. Disconnect tubes from the secretion jar lid.
3. Unhinge full secretion jar from the secretion jar mount and dispose of secretion fluids according to national
disposal regulations.
4. Procure used secretion jars to the reprocessing cycle.
5. Connect suspended tubes with new, ready to use secretion jar.
7.7

Function control

To obtain trouble-free operation of the suction pump all the components and functions of the pump system have to
be tested prior every assignment.
General functions:
1. Use the power switch "I/O", at the rear of the device, to switch the device on, indicator light, LED, is illuminated.
2. Neither the Infiltration pump nor the suction pump or the motor of the Conform Cannula Handpiece may be
actuated without the pedal beeing pressed.
Suction pump:
1. Press foot switch "Vacuum" (blue) at the Vario pedal to activiate the suction pump.
2. Hold hand in front of the "Exhaust" (Rear of the device). Airflow is perceptible.
3. Turn Vacuum Controller (Vacuum) clockwise, equivalent to vacuum maximum.
4. Control vacuum performance at the openings of the cannula. Suction performance is high.
5. Turn Vacuum Controller anticlockwise all the way to the stop, equivalent to vacuum minimum.
6. Control vacuum intensity at the openings of the cannula. Suction performance is low.
7. Use the power switch "I/O", at the rear of the device, to switch off device, indicator light, LED, goes off.
Infiltration pump:
1. Turn Infiltration Controller, "Infiltration", clockwise, all the way it goes, equivalent to maximal flow rate.
2. Press pedal on the Vario Pedal, infiltration pump starts.
3. Turn Infiltration Controller couterclockwise, all the way it goes, equivalent to minimum flow rate.
4. Infiltration performance is reduced up to a stand still.
5. Once the maximum infiltration amount is set, the delivery of infiltration liquid can be regulated by pedal.
Conform Cannula Handpiece with Electronicmotor:
By pressing the foot switch "Motor" (yellow) the electronic motor of the Conform Cannula Handpiece is activated
and runs with a constant speed. Vibrations at the Conform Cannula Handpiece are perceptible.
Malfunctions and troubleshooting:
To solve problems refer to chapter 10 "Malfunctions and troubleshooting".