Vacuson 60 LP
9.7
Seals and tubes
To insure proper function of the Suction Pump, all tubes and seals must be periodically checked,
and replaced after at most 250 sterilization cycles or five years of operation. Only when in per-
fect order can sufficient vacuum be built up.
Defective seals and tubes must be replaced.
1
1.
Seal (O-Ring) at intake nozzle of control unit for connection with bacteria filter ----------- 2 units ------------ REF 4063
2.
Connecting tube between bacteria filter and secretion jar lid (VACUUM) -------------------- 1 unit -------------- REF 4155
3.
Seal (O-Ring) at connection nozzle (unscrewable, VACUUM) at secretion jar lid ----------- 4 units ---------- REF 4064
4. Seal (O-Ring) at overflow protection system at jar lid ---------------------------------------------- 1 unit ------------ REF 28958
5.
Main seal between jar lid and jar ------------------------------------------------------------------------ 1 unit ------------ REF 28957
6.
Seal (O-Ring) at connection nozzle (unscrewable, PATIENT) of secretion jar lid ------------ 4 units ----------- REF 4063
7.
Suction tube between connection nozzle (PATIENT) and cannula ------------------------------ 1 unit ------------- REF 6026
Operating Instructions, Vacuson 60 LP, REF 31793, V34/17