7.4

Adjusting the programs

Values for operation settings depend on the connected handpiece or contra angle as well as the task to be
performed.
7.4.1

Step 1: Selecting the handpiece or transmission ratio

Depending on the handpiece or contra angle attached to the motor, the corresponding transmission ratio
has to be adapted accordingly by using the "HP" or the "PR" key.
Press the "HP" key to select Program 1 with the 1:1 contra angle
Press the "PR" key to select Program 2 or 3. The desired contra angle has to be selected in the configuration
menu before you can use it.
Table of possible contra angles
Name of the
handpieces/contra angles
with transmission ratio
Drill handpiece, 1:1
Drill contra angle, 16:1
Drill contra angle, 20:1
Drill contra angle, 32:1
Drill contra angle, 70:1
* The limitation of the AS-range (Automatic Stopper) can be adjusted in the Configuration Menu.
Handpieces or contra angles that don't belong to one's own assortment can be deactivated in the Configuration
Menu. That makes finding the right handpiece in the handpiece list easier, because the list of available handpieces
becomes shorter.
12
Speed
Speed
Display
min. rpm
max. rpm
1 : 1
300
16 : 1
20
20 : 1
15
32 : 1
10
70 : 1
5
Torque
Torque
min. Ncm
max. Ncm
40,000
1
6
2400
5
27
1700
10
70
1000
10
55
600
10
55
MD 11 Operating Instructions, Art. No. 31665, V10/14
MD 11
Limit
AS-range
AS-range
(Factory def.)
rpm
rpm (*)
No
up to 20
20 – 45 *
up to 20
15 – 45 *
up to 20
10 – 45 *
up to 20
5 – 45 *