Manual cleaning and wipe disinfection

Phase
Step
I
Cleaning
II
Drying
III
Wipe disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
RT:
Room temperature
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. To avoid the risk of corrosion/malfunctioning,
allow any fluid that has entered the product to drain out immediately.
Phase I
Clean the product under running tap water, using a suitable cleaning
brush until all visible residues have been removed from the surfaces.
Move non-rigid components (such as unlocking buttons, levers, sliders)
during cleaning.
Phase II
Dry the product in the drying phase with suitable equipment (e.g. cloth,
compressed air).
Phase III
Wipe all surfaces of the product with a single-use disinfecting wipe.
7.8

Mechanical cleaning/disinfection with manual pre-cleaning

Note
The cleaning and disinfecting machine must be of tested and approved
effectiveness
(e.g.
FDA
DIN EN ISO 15883).
Manual pre-cleaning with a brush
Phase
Step
I
Brushes
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. To avoid the risk of corrosion/malfunctioning,
allow any fluid that has entered the product to drain out immediately.
Phase I
If necessary, clean the product with a suitable cleaning brush until all
discernible residues have been removed.
Move non-rigid components (such as unlocking buttons, levers, sliders)
during cleaning.
8
T [°C/°F]
t [min]
RT (cold)
-
RT
-
-
>1
RT (cold)
0.5
RT
-
approval
or
CE
mark
according
T [°C/°F]
t [min]
RT (cold)
RT (cold)
1
Conc. [%] Water quality
-
D–W
-
-
-
-
-
FD–W
-
-
Phase IV
After the specified exposure time (at least 1 min), rinse the disinfected
surfaces under running FD water.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable equipment (e.g. cloth,
compressed air).
After manual cleaning/disinfection, check visible surfaces for residue
and repeat the cleaning/disinfection process as required.
Note
The cleaning and disinfection machine used for processing must be serviced
to
and checked at regular intervals.
Conc. [%]
Water quality
D–W
D–W
Phase II
Rinse the entire product (all accessible surfaces) under running water.
Move non-rigid components (such as unlocking buttons, levers, sliders)
during cleaning.
Chemicals
-
-