Speichern von Messdaten
Beispiel, um DATE zu speichern (Datum und Uhrzeit speichern): ON, SET (Messbedingungen) ON,
PARA (Messparameter): ON, DELIM (Trennzeichen): , (Komma), QUOTE (Anführungszeichen) "
(doppelte Anführungszeichen)
"HIOKI E.E. CORPORATION","IM3536","Ver. 1.00",
"Serial No. 123456789"
"DATE","11-11-30"
"TIME","10:10:06"
"FREQ","1.0000E+03","Hz"
"V","1.000","V"
"LIMIT","OFF"
"RANGE","AUTO","10k","ohm"
"LOW Z","OFF"
"JUDGE SYNC","OFF"
"JUDGE","OFF"
Titel
"SPEED","MED"
"TRIG","INT"
"AVG","OFF"
"DELAY","0.0000","s"
"TRIG SYNC","OFF"
"DCBIAS","OFF"
"OPEN","OFF"
"SHORT","OFF"
"LOAD","OFF"
"CABLE","0","m"
"SCALE","OFF"
"Z[ohm]","OFF","PHASE[deg]","OFF"
"4.983329E+03","","0.074",""
Beispiel: Parameter Nr. 1: Z (Impedanz [
Parameter Nr. 3: θ (Phasenwinkel der Impedanz [°]); Parameter Nr. 4: OFF
"4.983329E+03","","0.074",""
Messwert Parameter Nr. 1
Der Parameter Nr. 1 ist 4,983329 k
angezeigt für Parameter Nr. 4 oder Nr. 4, da sie auf OFF stehen.
142
CSV-Datei-Beispiel
Anführungszeichen
Trennzeichen
Messwerte interpretieren
Ω
]); Parameter Nr. 2: OFF;
Messwert Parameter
Nr. 2
Ω
und der Parameter Nr. 3 ist 0,074°. Keine Messwerte werden
","
Instrumentinfor-
mationen
Uhrzeit und
Datum
Messbedingungen
Messparameter
Mess-
Werte (siehe unten)
Messwert Parameter Nr. 3
Messwert Parameter Nr. 4