MDM300 & MDM300 I.S. Quick Start Guide
Determine the approximate
Typical MDM300 Sampling Arrangement
dew point by using the 'initial check'
procedure on the reverse
SAMPLE FLOW
1
Adjust the flow rate
using the
2
outlet flow control
System
at Atmospheric pressure
valve to
between
0.2 and 0.5 l/min
Atmospheric
3
Turn the instrument
on. The status display
will read
Adjust the flow rate
'Initializing
inlet flow control valve
internal sensor'
between
0.2 and 0.5 l/min
4
Turn the instrument on.
The status display will read
'Initializing internal sensor'.
The instrument will begin
Measurement mode. During the first few
minutes of this, the displayed moisture
value will drop rapidly below the actual
moisture value, before returning
VENT
to a stable reading.
1
Coalescing Filter
The instrument's display will show an initial
2
Particulate Filter
measured value, and the status display
will show 'Measurement in progress'.
3
Inlet Flow Control Valve
4
Outlet Flow Control Valve
Once 'Measurement in progress'
has disappeared, the instrument is close
to the final dew point.
Depending on the conditions, it may
continue to respond for a number of
minutes before tracking the measured dew point.
HIGH PRESSURE! High pressure gases are potentially hazardous. Energy stored in these
!
7
Michell Instruments
Connect the MDM to
the sample gas stream
Is the
equivalent
No
dew point
> - 40Cdp?
Do you
want to measure
Atmospheric
at Atmospheric pressure
Yes
or System
pressure?
Do you
System
want to measure
Adjust the flow rate
or System
using the
inlet flow control valve
pressure?
to
between
0.2 and 0.5 l/min
Turn the instrument
on. The status display
using the
will read
'Initializing
to
internal sensor'
After the initialization
period is complete,
close the outlet flow
control valve
4
and
fully open the inlet
flow control valve
3
.
Then adjust the flow
rate using the outlet
flow control valve
to between 0.2 and
0.5 l/min.
97244 Issue 02, July 2011
sue 0 , Ju y 0
s s ue 02, July 2011
Abb 6
2
4
8
Zubehörteile
INSTALLATION
3
5
6
11