Herunterladen Inhalt Inhalt Diese Seite drucken

Marzhauser Sensotech Profiler ST Betriebsanleitung Seite 180

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

All manuals and user guides at all-guides.com
Technical Data
Pin assignment for 1Vss measuring systems
1
2
3
4
14.4.2 ABS/SSI 1–3: D-Sub 9-Pin Plug
For connection of absolute measuring systems with the SSI
standard interface, use shielded twisted pair encoder cables
only. Do not route the encoder cables parallel to power cables.
Place the shield on both sides.
180
Pin
Signal
1
–U1
2
GND
3
–U2
4
ERR
5
–U0
6
+U1
7
VCC5
8
+U2
9
+U0
output signal: 1Vss, differential transmission, R
TTL input, pull-down resistor 1 kΩ  GND,
low pass 1 kΩ/10 nF
max. 500 mA, total load of all VCC5 ports max. 800 mA
differential transmission, R
Function
–sine
1
reference potential
–cosine
1
error
2
–reference mark measuring system
+sine
1
power supply 5 V
+cosine
1
+reference mark measuring system
L
= 120 Ω
L
Operating Manual
4
3
4
= 120 Ω

Werbung

Kapitel

Inhaltsverzeichnis
loading