Operation
5 Approach the machine zero for the axis to be corrected:
If the position display of the respective axis shows 0, the ma-
chine zero was successfully set/approached.
6 Position a measurement standard (e.g. glass scale) on your
7 Open the Position Correction 2/2 menu.
8 With
The number of the new segment appears in the Segment dis-
play field.
132
If you want to configure the position correction for an
axis with an absolute measuring system type SSI, di-
rectly approach the machine zero. For absolute measur-
ing systems, the machine zero is detected directly after
switching on the PROFILER ST.
If you want to configure the position correction for an
axis with an incremental measuring system type
1Vss or TTL (RS-422) and you have enabled the Set
machine zero with button function in the Origin
menu, position your measuring setup as desired. Set the
machine zero with
If you want to configure the position correction for an
axis with an incremental measuring system type
1Vss or TTL (RS-422) and you have enabled the Set
machine zero with reference mark function in the
Origin menu, approach the the desired reference mark
of the respective axis. To the left of the axis' position
display, the status display reference signal
from white to green.
Note: Refer to the description of the functions to deter-
mine the machine zero.
»
Chapter 6.7.3 Origin, page 140
measuring setup along the axis to be corrected.
add a new segment for the axis along which you
have positioned the measurement standard.
Note: Target and actual values for the 0 segment are pre-
defined and cannot be changed.
,
or
.
Operating Manual
changes