NOTICE
Loss of position correction values from deleting or over-
writing!
The Position Correction menu is not write-protected. If re-
quired, protect the Setup menu using a password against
unauthorised access.
»
X
1 Configure the following settings to determine the machine
2 In the Encoder menu, enable the check box REF for all axes
3 Save your entries with
4 Switch to the position display.
Chapter 6.7.2 System, page 138
Configuring the position correction
zero for all connected incremental measuring systems type
1Vss oder TTL (RS-422):
In the Origin menu, enable the desired function for each
axis to determine the machine zero.
»
Chapter 6.7.3 Origin, page 140
If needed, enter a value for the Offset from machine
zero on the affected axes in the Origin menu.
Save your entries with
Note: For absolute measuring systems (encoder type SSI)
the machine zero is set by hardware and not configurable.
whose machine zero you would like to set by reference
mark.
Note: The signals of reference marks from absolute mea-
suring systems (encoder type SSI) are not evaluated by the
PROFILER ST. The position correction is always done with
respect to the hardware set machine zero.
.
.