PROFILER ST
Remedy
Check whether the external power supply is properly connected to the
24V/DC port on the PROFILER ST and the socket.
Check the settings for the encoder type of the affected measuring system.
»
Chapter 6.7.6 Encoder, page 148
1 Disconnect the power supply of the PROFILER ST from the mains.
2 Check whether the affected measuring system is properly connected to the
PROFILER ST over the correct port.
3 Connect the power supply of the PROFILER ST to the mains.
4 Check the position display of the affected axis.
Disable the position displays of unused measuring channels.
»
Chapter 6.7.6 Encoder, page 148
Check the settings for the encoder type and the signal period/counting step of
the affected measuring system.
»
Chapter 6.7.6 Encoder, page 148
1 Check the position correction values of the affected measuring system.
2 Correct faulty position correction values or, if necessary, delete all position
correction values and configure the position correction once more.
»
Chapter 6.5.2 Configuring the Position Correction, page 130
Check the specifications of the affected measuring system.
»
Chapter 4.1 Intended Use, page 109
If Set machine zero with button function is selected: After switching on the
PROFILER ST, manually set the machine zero with
If Set machine zero with reference mark function is selected: Approach the
ref erence marks of all axes after switching on the PROFILER ST.
»
Chapter 6.7.3 Origin, page 140
Enable the position correction for the affected axis.
»
Chapter 6.7.6 Encoder, page 148
Troubleshooting
,