9 Move the measuring setup to the next desired measuring
10 Tap on the Target Pos. input field.
11 Enter the value of the measuring point approached on the
12 Confirm your entry with
13 With
NOTICE
Incorrect position correction from excessive deviation
between target value and actual value!
The Target Pos. and Act. Pos. values may deviate from each
other by a maximum of 127 µm. Check each value pair before
saving it.
14 Save the entered target and actual values for the segment
15 To define further segments (max. 128), repeat steps 8 to 14.
16 Follow steps 5–15 for all axes which are to be corrected.
Configuration of the position correction is now complete.
17 In the Encoder menu, enable the check box Corr for all axes
18 Save your entries with
You can now carry out corrected measurements.
point of the measurement standard.
measurement standard via the entry bar.
, adopt the currently measured position value in the
input field Act. Pos.
Note: If you would like to enter or modify the actual value
manually, tap on the Act. Pos. input field. Enter or modify
the actual value via the entry bar. Confirm your entry
with
.
with
.
Select the distance of the measuring points depending on
the travel range of your measuring setup (e.g. 10 mm).
for which you have set up the position correction.
.
.