Commissioning
NOTICE
Possible equipment damage from connecting inappropri-
ate measuring systems or using wrong encoder cables/
cable adapters!
Only connect incremental measuring systems type 1Vss or TTL
(RS-422) without external power supply to the Encoder 1–3
ports.
Only connect absolute measuring systems type SSI without
external power supply to the ABS/SSI 1–3 ports.
Use appropriate encoder cables and cable adapters only.
»
5 Connect your incremental measuring systems to the
6 Connect your absolute measuring systems to the
WARNING
Possible electric shock and equipment damage from con-
necting foreign or damaged power supplies!
Only use the delivered 24 V/30 W power supply. Immediately
replace the power supply if there is damage on the housing or
cable of the power supply. Only use replacement components
from Märzhäuser Sensotech.
Possible electric shock and equipment damage from con-
necting to non-specified mains!
Ensure that your mains voltage matches the label on the power
supply before connecting the power supply to the mains.
7 Connect the power supply cable to the 24V/DC port.
8 Connect the power supply to the mains.
9 Switch on the PROFILER ST (ON-OFF switch to position I).
114
Chapter 14.4 Ports, page 179
Encoder 1–3 ports of the PROFILER ST.
ABS/SSI 1–3 ports of the PROFILER ST.