PROFILER ST

14.2 Encoder Interface

1
2
Security functions
Processor monitoring
Voltage monitoring
Surge/short-circuit
protection
(external voltages)
Parameter
Position signal
Reference signal
evaluation
Error signal evalua-
tion
Max. input frequency
Max. interpolation
factor
Input resistor
Supply voltage meas-
uring system
differential transmission
sampling frequency interpolation
reset in case of undervoltage,
logic voltage (3.3 V)
measurement of internal voltages
(3.3 V, 5 V, 24 V/DC)
VCC5 + VCCEXT, shutdown is
automatically reset after the end
of the overload condition
TTL
(RS-422)
standard
1
yes
on
request
25 MHz
3
4
5
120 Ω
5 V
6
Technical Data
1Vss
0,6...1,2 Vss
1
yes
on
request
25 MHz /
22 kHz
2
12.868
4
120 Ω
5 V
6