Switching The Unit On And Off - Sartorius Stedim Biotech Sartocheck 4 plus Betriebsanleitung

Bag tester / bag tester multiunit
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

7.2 Switching the Unit
ON and OFF
Fig. 7-3 | Login menu
Fig. 7-4 | Main Menu
28
Operation
Downloaded from
www.Manualslib.com
If you want to switch the unit on, proceed as follows:
Before you switch on the power supply and the compressed air, compare the
connection data for the unit (see the Chapter 01 Technical Specifications) with
the data for the mains current and the compressed air supply.
Connect Sartocheck
cord into the wall outlet, install all tubing connections. The unit is now ready for
operation.
Login menu
User ID:
Password:
OK:
Cancel:
Important Note:
Name:
Password:
Operators must log-in before activating the Main Menu.
The operator logs in by pressing the "Login" key on the keypad. After pressing the
"Login" key, the user will be asked for their name and password. When both are
entered correctly, the keypad and touch screen are released for use. When the opera-
tor stops working with Sartocheck
pressing the "Logout" key. Sartocheck
operation until another operator logs in.
Information line with model, version, system number and user name.
Menu display: Main Menu
F1-F8: Select the function menus with the function keys
Selects the function menus
Functional unit selected and operational status.
The Main Menu appears.
t
Before starting a test, allow the unit to warm up for about 20–30 min.
For shut-down times < 30 min and constant ambient temperatures, the warm-up
phase can be shortened accordingly. For more information about the tests, see
Chapter 7.3 ff.
t
Program the data for the test routine.
t
The Sartocheck
t
Start the test.
Only service technicians or trained personal may install connections and start up
the unit!
manuals search engine
®
4 plus Bag tester to the mains supply by plugging the power
Enter operator's name
Enter operator's password
Confirm correct entry
Delete entry
®
When Sartocheck
4 plus Bag tester is delivered it is factory set
for the following user:
mentor
mentor
®
4 plus Bag tester, they must always log out by
®
4 plus Bag tester is then blocked for further
®
4 plus Bag tester ready for operation (by the logged-in user).