7.14 Installed bag Test

Fig. 7-30 | Test parameters for installed bag test
Downloaded from
www.Manualslib.com
This test sequence is designed for leak testing of 3 dimensional bags (e.g. STR
Cultibags) being installed in the appropriate STR holder equipped with fleeces.
The test sequence is based on the change in pressure within the bag system including
connected tubes and vent filters.
The pressure drop measured in the process can be used as a direct measure of the
bag set-up leak tightness and absence of flaws or similar defects, provided that the
upstream volume of the system is constant.
Enter parameters for the plate holder test
Test pressure:
Stabilization time:
Test time:
Max. pressure drop:
Filling time:
Please verify the test parameters for the bag being tested. Applying too high a test
pressure may jeopardize the integrity of the bag.
Since the viscosity of remaining liquid may influence the leak rate of an eventual
flaw, please do not mix up pre-use test values with specifically validated post-use
values.
This test can also be used for testing vessels as long as they resist the applied
pressure.
Enter record data 1 (up to 32 characters per field)
Company:
Building:
Department:
Manufact. site:
Product:
Product lot:
Bag type:
Bag lot:
Locks
After entering with the Shift + F keys, the programmer can lock the data to prevent
any changes. The dialog boxes are then highlighted in gray. The user can no longer
change the entered data. The relevant record data are entered in the menu
"Test program: Record data 1."
Entry of record data 2
Nom. bag volume:
Test gas:
Comment:
Comment 2:
Comment 3:
manuals search engine
from 3–50 mbar
1-1080 minutes.
1-1080 minutes
1-programmed test pressure
1-1080 minutes
Company name
Building name; house number
Identification of the department
Production sector
Product name
Batch number
E.g. STR Cultibag 1000L
Number of the bag batch
Nominal volume of the bag
Type of gas
Name of operational process
Enter test-relevant remarks
Enter test-relevant remarks
Operation
41