Herunterladen Inhalt Inhalt Diese Seite drucken

Sartorius Stedim Biotech Sartocheck 4 plus Betriebsanleitung Seite 60

Bag tester / bag tester multiunit
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Fig. 7-78 | Service Menu: Software configuration
60
Operation
Downloaded from
www.Manualslib.com
Edit Parameters
F1:
Password
You can Activate|Deactivate the password expiration date. When the password
expiration date is activated, the password must be renewed after the password
expiration date. Minimum duration is 1 day and a maximum 366 days. After
the password expiration date has expired, you will automatically be prompted
to renew the password. The max wrong password attemps can be set from
1-999. The minimum password length can be set from 1-999. Change pass-
word new user: if activated, the newly created user must change the password
during the initial connection.
F2:
Test program
Automatic Test Time Mode can shorten the test time if the values are
constant within a defined tolerance. When Automatic Test Time Mode is
deactivated, the test time is used that is set during the programming of the
test program. The sequential tests function is used for repeating the same test.
F3:
Hardware
Furthermore, the internal printer can be turned ON |OFF.
Management unit parameters
– Save parameters to SD card, F4
– Load parameters from SD card, F5
With these two options, you can save Sartocheck
unit parameters to an SD card, and | or load saved data from an SD card into
Sartocheck
Sartocheck
F4:
Auto Log Off
To activate |deactivate the auto log off and to define the timeout
F5:
Results database
To define after how many test results the Sartocheck
give a warning message that the memory will soon be full
F6:
PDF
To activate | deactivate saving of the test result files in PDF format.
To select the PDF page format A4 | Letter
F7:
Data input
Input device: Keyboard/Keyboard+barcode/Only barcode scanner
Report field: select which filed should be scanned
for recognition by the barcode
manuals search engine
®
4 plus Bag tester. The management unit parameters stored in
®
4 plus Bag tester are overwritten with those on the SD card.
®
4 plus management
®
4 plus Bag tester will

Werbung

Kapitel

Inhaltsverzeichnis
loading

Verwandte Produkte für Sartorius Stedim Biotech Sartocheck 4 plus