7.19.3.2 Calibration and
Adjustment of the External
Pressure Sensor Input
Fig. 7-89 | External reference
Fig. 7-90 | External pressure sensor input
Fig. 7-91 | Result calibration
Fig. 7-92 | Result adjustment
Downloaded from
www.Manualslib.com
The calibration of the external pressur sensor
Filter tester. Connect the Sartocheck
air supply. Connect the electrical outlet "Ext. Sensors" to the external reference
current source (4-20 mA). Enter the data on your reference into the designated list
(see Figure 7.34).
Service-Menu Calibration | Adjustment
F1:
Current Reference
F4:
F5:
F6:
F7:
F8:
With Page Down, start the actual calibration and | or adjustment.
Enter the corresponding target value of the reference current supplied and wait
for approx. 3 s. If there is a difference between the predefined value and the value
read by the reference, then enter the read-in value in the table. If the values match
exactly, go with PAGE DOWN to the next reference point. Continue to repeat this
procedure until the last reference point has been run up and processed. Calculation
of the data is now complete. Depending on the procedure selected (calibration or
adjustment), the results table is displayed.
During calibration, the results are displayed as a deviation of the adjustment curves.
This classified result is output in addition to the table providing the values for
the current supplied. Use "Print" to generate a printout for the calibration just
carried out.
During adjustment, a new adjustment curve is calculated and activated using
the entered values. The data of the adjustment can be printed out on the internal
printer after the adjustment is completed. The new adjustment curve can then
be checked with a calibration.
"Page up" takes you back to the selection screen for test units (MultiUnits) and
sensors. "Page down" is used to immediately start an additional calibration or
adjustment, depending on "F10 perform calibration | adjustment?".
manuals search engine
®
select Calibration | Adjustment
Model name
Serial No.
Certificate No.
Internal No
Comments
input
is only valid for the Sartocheck
4 plus to the current and compressed
Operation
®
4 plus
65