Fig. 7-8 | Test parameters: External connections
Fig. 7-9 | Test parameters: Diffusion test
32
Operation
Downloaded from
www.Manualslib.com
In the menu "Test parameters: External connections," decisions about external
connections and the configuration are entered. The function keys are used here to
enter the appropriate settings. The following external connections and parameters
must be configured for all tests:
– Standard test (Yes|No)
– External pressure sensor (Yes | No)
– External valves (Yes | No)
The parameters for the standard test are defined in the configured Service Menu.
Unlike the defaults (standard values), the values can be individually adjusted.
For the diffusion test, you have the following options:
Pressure unit:
Automatic test mode: No|Yes
Enter parameters for the diffusion test
Test pressure:
Stabilization time:
Test time:
Max. diff.:
Net volume:
In the menu "Test parameters: Diffusion test," the parameters for the diffusion test
are entered with or without net volume measurement.
For tests with volume measurements:
– Test pressure
– Stabilization time
– Test time
– Max. diffusion
When conducting tests without volume measurement, net volume should
be entered separately.
The test parameter data can be found in the validation documents of the respective
filter cartridge.
Locks
The locks are activated for the test parameters (default). After entering with the
Shift + F keys, the programmer can lock the data to prevent any changes.
The dialog boxes are then highlighted in gray. The user can no longer change
the entered data.
manuals search engine
mbar|psi | hPa
from 100 – 8000 mbar
in min.
in min.
in ml/min
in ml