Fig. 7-87 | Result of the calibration
Fig. 7-88 | Result of the adjustment
64
Operation
Downloaded from
www.Manualslib.com
During calibration, the results are displayed as a deviation from the adjustment
curves. If the deviation is below 0.05 % (0.3 mbar) FS for the Bag tester or 0.1 %
(9.5 mbar) FS for the Filter tester, an adjustment is not required.
During adjustment, a new adjustment curve is calculated and activated using
the entered values. The data of the adjustment can be printed out on the internal
printer after the adjustment is completed. The new adjustment curve can then
be checked with a calibration.
"Page up" takes you back to the selection screen for test units (MultiUnits) and
sensors. Depending on "F10 perform calibration | adjustment?", "Page down" is used
to immediately start an additional calibration or adjustment.